Features
ETL Listed to UL 1331 and CSA Z9170-1
Conforms to NFPA 99 and CGA standards
Accepts DISS gas specific nut & nipple sets
Ease of maintenance without removing the faceplate using a cartridge insertion tool.
Pin indexed faceplate and back
Removable front assembly
All outlets are 100% tested for flow and leaks
360 swivel inlet tube
Metal back and nameplate
Provides support for dispensing equipment without the use of additional brackets
Made in the U.S.A.
DISS Style Console Outlets Medical Vacuum Male DISS Style Outlet
The Frontall DISS medical gas console/headwall outlets represent the most technically advanced means of providing medical gas connections to the central piping system. The outlets incorporate a unique front-loaded valve cartridge feature which makes the outlets easy to repair, saving valuable time and money for the health care facility.
Modular single outlets incorporate nameplate color-coding in accordance with NFPA 99
Dispense equipment without the use of slide brackets
Perform routine maintenance without removing nameplate and/or gas specific portions
Primary check valve can be removed for service without shutting off gas supply
Secondary check valve operates automatically to stop the free flow of pressure gas
No secondary check for the vacuum or WAGD services
Double O-ring seal between nameplate and rough-in is serviceable without shutting off gas service, supports 10 lbs. at 2? from nameplate without leaking
Scratch-resistant, durable, color-coded nameplate
Rough-in supply tube is 7? in length of ? O.D. copper Type K for all gas services
Clean for oxygen service in accordance with current CGA Pamphlet G-4.1
Test before use in conformance with NFPA 99, local and state codes