Features
C-arm unit also includes table-side control boxes and control cabinet.
GE Omega IV angio table with weight limit up to 450 pounds, features ?floating top? for flexible patient positioning.
Table top height adjustment measures 30.5 to 42.5 inches.
Table Side User Interface (TSUI): Smart box or Smart handle.
Table Side Status Control (TSSC) and Innova central touchscreen.
Innova Console includes the following components:
Innova Digital Color LCD Monitor
Exposure Hand Switch
Innova Console
Console keyboard and mouse
X-ray high-frequency and high-voltage generator.
X-ray tube and collimator.
Image detection unit.
In-room and control monitors.
DICOM worklist compatibility.
External chiller provides continuous water cooling.
Ceiling and table-mounted radiation protection.
Accessories: table panning device, arm support, arm board, dual footswitch with table unlock, IV pole and mount, Velcro quick strap, tilting table, footrest/shoulder rests ? ask your Avante representative for more information.
Available Options
Injector interface
Table head extender
Diamentor dose monitoring
Innova IVUS
Downscan
Innova central touch screen
Innova IQ table
UPS
Wall-mount suspension
In-room 3D
CRT monitor suspension
TSUI options: Smart handle, smart box, second smart box, second Smart handle, second TSSC, remote TSSC, remote smart handle, remote smart box
OMEGA table foot end rail
Second footswitch with cover
Workstation (including power supply, imaging processing system, TV display system)
Remote Stand
Multiple software options ? ask your Avante representative for more information
GE Innova 2100 Cath/Angio System
Complete cardiovascular imaging system delivers advanced image quality, patient positioning
The GE Innova 2100 is a complete cath/angio system that offers clinicians easy patient positioning, advanced image quality, and innovative dose management. The Innova 2100 provides consistently high imaging performance while allowing clinicians to complete interventional and diagnostic angiography procedures.
Generator
Nominal output power: 100 kW
Max kW of generator?s output when loading time is 0.1s, 100kW
Range of frequency used by the high-voltage generator: 20 kHz – 60 kHz
kV-mA range: kV from 40 kV to 125 kV | Maximum mA: 1000 mA
Large focus: Maximum mA: 1000 mA | Maximum power: 100 kW
Small focus: Maximum mA: 400 mA | Maximum power: 48 kW
Compressed focus: Maximum mA: 200 mA | Maximum power: 16 kW
Gantry
L-arm rotates on its vertical axis +/- 100? (+/- 95? motorized limit set)
Offset C-arm (pivot) permits -117?/+105? RAO/LAO rotation
The C-arm permits 50? cranial and 45? caudal angulation of the imaging system
SID Fully motorized (8.9 cm/s – 3.5 in/s)
SID range of 85 cm to 119 cm (33.5 to 46.8 in) for Innova 2100-IQ. Travel distance of 34 cm (13.4 in)
X-ray tube Focal Spot to Isocenter 72 cm (28.3 in)
Isocenter to the floor of 107 cm (42 in)
Omega IV Table
Tabletop filtration: Less than 0.85 mm aluminum equivalent
Mattress: 2.5 cm (1 in) mattress + slicker: less than 0.70 mm aluminum equivalent
Maximum patient weight: 204 kg (450 lb)
Maximum table load: 204 kg (450 lb)
Tabletop length: 300 cm (118 in)
Tabletop width: 48 cm (18 in) in Patient Trunk Area – 67 cm (26.4 in) max
Horizontal float movement: 8-way
Horizontal motorized movement: N/A
Longitudinal speed: N/A
Longitudinal travel: 110 cm (43.3 in)
Equivalent fluoroscopic coverage: 126 cm (49.6 in)
Transverse travel: manual +/- 14 cm (+/- 5.5 in)
Vertical travel: 30 cm (12 in)
Vertical travel above floor: 2 cm/s (0.8 in/s) at 50 Hz | 2.5 cm/s (1 in/s) at 60 Hz
Table base: 61 cm x 52 cm (24 in x 20.5 in)
Table rotation: +/- 180?
Innova Digital
DICOM V3.0 implementation
Store up to 4,000 sequences and photos
Innova Imaging
Available input screen size of detector (cm): 0 Mag – 20 | 1 Mag – 17 | 2 Mag – 15 | 3 Mag – 12
Horizontal center resolution (not less than); (lp/mm): 0 Mag – 2.5 | 1 Mag – 2.5 | 2 Mag – 2.5 | 3 Mag – 2.5
Hi-resolution 1249 line analog video output
Hi-resolution 1249 line motor, 50 Hz
Visibility range: 4 cm – 13 cm (using 20 cm PMMA and 16 cm PMMA step wedge)