Features
New – Masimo SET? Pulse Oximetry ? ask your Avante Representative for details
Optional Integrated Dr?ger Anesthesia Gas Bench expands measurement capabilities
Optional agent bench is compact and part of main unit. All in one monitor!
Large (15″), touch-screen with high-resolution flat panel display
Displays as many as eight waveforms
Multi-lead simultaneous ECG monitoring
Automatically set alarm limits
Graphical and tabular trending
Color-coded alarms
Patient data entry
Networking capability
Works with EMR and wireless telemetry systems
Large time stamp in upper right corner
Standby button
Volume and Sounds are adjustable and can be turned off completely
Option to print Numerical Data Only
Color of the waveforms can be changed
Battery backup (two removable batteries)
Available with a built-in thermal printer
Waveline Pro Touch-Screen Anesthesia Monitor
Superior performance with a high-resolution touch-screen
The Waveline Pro displays as many as eight colorful waveforms on a crisp, high-resolution 15-inch touch-screen that optimizes the speed of patient care. The user can make quick screen adjustments, set default settings, alarm limits, and manage up to 72 hours of detailed patient data. Designed for a fast paced work environment, the Waveline Pro patient monitor offers an extremely simple and adaptable user interface. Patient information along with vital-signs settings can be quickly modified to meet the needs of a patient’s changing condition, converting from a basic vital signs monitor, to a continuous bed side monitor, to an operating room monitor. Now with Masimo SET? Pulse Oximetry!
This patient monitor offers standard measurements of: non-invasive blood pressure, 3-lead or 5-lead ECG with arrhythmia detection, motion tolerant SpO2, Temperature, and Respiration rate. End-tidal CO2, Anesthetic Agent measurement, Cardiac Output and Invasive blood pressure are available options.
The Waveline Pro monitor offers several connective solutions to network multiple monitors and/or manage patient data on an electronic medical records platform or a HL7-based hospital information system. It offers Ethernet and RS-232 connections with an open source communication protocol, and it is HL7 compliant. The HL7 network protocol will allow for all patient information and vital sign trends to be transferred and stored on a hospital information system.
Performance
Display: 15-inch color touch-screen
Trace: 8 waveforms
Indicator:
Alarm indicator
Power indicator
QRS beep and alarm sound
Trend time: 1 – 72 hour
Recorder:
Built-in, thermal array, 3 channels
Record width: 48mm
Recorder paper: 50mm
Record speed: 25mm/s, 50mm/s
ECG
Input: 5-lead ECG cable and standard AAMI line for connection
Lead Choice: I, II, III, aVR, aVF, aVL, V, V1-V6, TEST
Gain Choice : x0.5, x1, x2, x4
Frequency Characteristic: 0.05 ~ 35 HZ (+3dB)
ECG Waveforms: 7 channels
Penetration Voltage: 4000VAC 50/60Hz
Sweep Speed: 12.5, 25, 50 and 100 mm/sec (left to right or right to left)
HR Display Range: 30 ~ 300bpm
Accuracy: ?1bpm or ?1%, whichever is greater
Alarm Limit Range Setting:
Upper limit 100 ~ 200bpm,
Uower limit 30 ~ 100bpm
RESP
Measure Method: RA-LL impedance
Range: 0 ~ 120 rpm
Accuracy: ?3 rpm
Alarm Limit Setting:
Upper limit 6 ~ 120 rpm,
Lower limit 3 ~ 120 rpm
Sweep Speed: 12.5, 25, 50 and 100 mm/sec (left to right or right to left)
NIBP
Measuring Technology: automatic oscillating measurement
Cuff Inflating: <30s (0 ~ 300 mmHg, standard adult cuff)
Measuring Period: AVE0.02 % Pulse Amplitude and % Transmission >5
SpO2: ?2 %
PR: ?3 bpm
Modes:
Averaging mode: 2,4,8,10,12,14 and 16 s
Sensitivity: Normal, APOD and Maximum
PR Accuracy (non-motion):
Neonate: 70?100 %: ? 3 %, 0?69 % : unspecified
Adult Pediatric Neonate: 25?240 bpm: ?3 bpm
PR Accuracy (motion):
Adult Pediatric Neonate: 25?240 bpm: ?5 bpm
Measuring Range:
SpO2: 1?100 %
PR: 25?240 bpm
Perfusion: 0.02?20 %
IBP
Measurement Range: -50 ~ 300mmHg
Channel: 2 channels
Pressure Transducer: sensitivity, 5µV/V/mmHg
Impedance Range: 300 ~ 3000?
Transducer Sites: ART, PA,CVP, RAP, LAP, ICP
Unit: mmHg/kPa selectable
Resolution: 1mmHg
Accurancy: ?1mmHg or ?2%,whichever is greater
AlarmRange: -10 ~ 300mmHg
EtCO2 (Optional)
CO2 Measurement Range: 0 ~ 99mmHg
Accuracy:
?2mmHg (0 ~ 38mmHg)
39-99mmHg ?5% of reading +0.08% for every 1mmHg (above 38mmHg)
Sampling Rate: 50 ml/min
Initialization Time: 30 seconds (typical), reaches ?5% steady-state accuracy within 3 minutes.
Respiration Rate: 0 ~ 150 breaths/min
Mode: adult, neonate
Cardiac Output (Optional)
Measurement Method: Thermodilution Method
CO2 Measurement Range:
C.O.: 0.1 to 20 L/min
TB: 23 to 43
TI: 0 to 27
Resolution:
C.O.: ?5% or ?0.1 L/min, whichever is greater, as measured using electornically generated flow curves.
TB, TI: 0.1
Accuracy:
?5 or ?0.1 L/min, whichever is greater, as measured using electronically generated flow curves.
TB, TI ?0.1 (without sensor)
Alarm Range: TB 23 to 43
Repeatability: C.O. ?2% or 0.1 L/min, whichever is greater, as measured using electronically generated flow curves.
Anesthetic Agents (Optional)
Method: Infrared absorption
Gas Sorts: Halothane, Isoflurane, Enflurane, Sevoflurane, Desflurane, CO2, N2O, O2 (optional Automatic Agent ID)
Measurement Range:
Halothane, Isoflurane: 0 – 8.5%
Enflurane, Sevoflurane: 0 – 10%
Desflurane: 0 – 20%
CO2: 0 – 10%
N2O: 0 – 100%
O2: 0 – 100%
Bias:
Halothane, Isoflurane, Enflurane, Sevoflurane, Desflurane: ?(0.15 Vol% + 15% rel.)
CO2: ?(0.5 Vol% + 12% rel.)
N2O: ? (2 Vol% + 8% rel.)
O2: ?3 Vol%
Networking
Industry standard 802.11b/g wireless network
Power
Source: External AC power or internal battery
AC Power: 100 – 240VAC, 50/60Hz, 150VA
Battery: Built-in and rechargeable lithium ion
Operating TIme: 3+ hours
Environmental
Temperature:
Operating: 5? – 40? C
Storage: -20? – 65? C
Humidity Range:
Operating: ?;80 %
Storage: ?80 %
Other Standard Feature
OxyCRG, drug dose calculation, cascading ECG, On screen NIPB trends (up to 250 readings), user set defaults, Arrhythmia detection, ST segment